Pro-MAM Performance

03-601

The phantom designed to be used for routine quick and ongoing checks of imaging performance of mammography systems. It allows evaluating particularly those aspects which are most liable to deterioration.

Technical data (can be modified to customer specifications):

  • 240 mm diameter semicircle
  • first half of the test module contains:
    - 6 groups of 10 mm long multi-directional filaments of the following diameters: 0.40, 0.35, 0.30, 0.25, 0.225, 0.20 mm
    - 6 groups of microcalcifications in the following ranges: 354-224, 283-180, 226-150, 177-106, 141-90, 106-63 μm
    - 6 groups of 3 low contrast details, 3mm in diameter, producing nominal contrast at 28 kVp of: 0.04, 0.03, 0.02, 0.015, 0.01, 0.005 (each contrast is presented 3 times)
  • second half of the test module contains:
    - clinically realistic breast tissue (not tissue equivalent to but simulating the appearance of the breast tissue) containing 6 groups of low contrast details with the same specifications as in the other half
  • 6 PMMA plates 10 mm thick
  • 2 PMMA plates 5 mm thick
  • carrying case

Product features:

  • complies with:
    – IEC 61223-3-2
  • this is not an ACR approved phantom for accreditation
  • CE certified
  • the manual provides detailed guidelines for carrying out each test, results assessment and registration

Ask a Distributor for an offer

Locate a local distributor and access exclusive medical offers tailored to your needs.

Share this product
Close Cookie Preference Manager
Cookie Settings
By clicking “Accept All Cookies”, you agree to the storing of cookies on your device to enhance site navigation, analyze site usage and assist in our marketing efforts. More info
Strictly Necessary (Always Active)
Cookies required to enable basic website functionality.
Made by Flinch 77
Oops! Something went wrong while submitting the form.