The compact and easy to use phantom for performance evaluation of fluoroscopic systems according to Center for Devices and Radiological Health (CDRH) specifications. It also meets recommendations of AAPM Report No. 60 “Instrumentation Requirements of Diagnostic Radiological Physics”. It is optimized for both under- and over-table fluoroscopic tubes.
Technical data (can be modified to customer specifications):
set of acrylic plates making total thickness of 193 mm
thanks to modular construction different total thicknesses can be easily set up
size of acrylic plates is 177.8 x 177.8 mm
2x 2.3 mm aluminium filters can be screwed underneath the acrylic plates
four beads embedded on the top plate can be used as orientation points for collimation setup
phantom stands on two legs 100 mm above tabletop
one leg is a probe holder
back plate with a handle can be easily unscrewed for over-table measurements
additional 1.6 mm copper filter simulates the presence of a 2 mm thick layer of barium sulfate, and can be used to measure the air kerma rate (free in air)
3.2 mm lead plate simulates maximum attenuation, and can be used to measure the maximum air kerma rate (free in air)
two types of a fluoroscopic image quality test object containing 8 low-contrast holes in an aluminium disc and 8 high contrast meshes or a high contrast resolution lead plate (from 0.6 to 5.0 LP/mm).
heavy duty carrying case
Under-table set up
Over-table set up
Product features:
complies with: – Nationwide Evaluation of X-ray Trends (NEXT) Protocol for 2003 Survey of Fluoroscopic X-Ray Systems – IEC 61223-3-1 – AAPM Report No. 60 “Instrumentation Requirements of Diagnostic Radiological Physics”
CE certified
the manual provides detailed guidelines for carrying out each test, results assessment and registrationt
Ask a Distributor for an offer
Locate a local distributor and access exclusive medical offers tailored to your needs.
By clicking “Accept All Cookies”, you agree to the storing of cookies on your device to enhance site navigation, analyze site usage and assist in our marketing efforts. More info
By clicking “Accept All Cookies”, you agree to the storing of cookies on your device to enhance site navigation, analyze site usage and assist in our marketing efforts. More info
Oops! Something went wrong while submitting the form.
Informacja dla klientów
Zgodnie z obowiązującymi przepisami prawa, niniejsza strona i prezentowane na niej treści są przeznaczone dla profesjonalistów związanych z wyrobami medycznymi i/lub ich obrotem. Zgodnie z powyższym, czy potwierdzasz, że jesteś profesjonalistą i chcesz zapoznać się z treścią strony?